The 5606 is just two inches (50 mm) in length with a sheath diameter of 1/8 inch (3.1 mm). Since it can be fully immersed over its entire temperature range, you don’t have to worry about calculating minimum immersion depth— immerse the entire probe, transition junction, and lead wires, in either non-corrosive liquids or dry mediums ranging from –200 °C to 160 °C.

The 100 Ω precision sensing element (alpha = 0.00385) is specially constructed to prevent the ingress of moisture while achieving calibration accuracies as good as ± 0.05 °C. You get both accuracy and flexibility built into a small, durable package.

The lead wires of the 5606 are made of single-conductor enameled copper wire which prevents moisture from wicking to the sensing element as it would if a more common, stranded lead wire were used. The diameter of the lead wire is no greater than 0.2 mm, making it easy to run lead wires through a freezer door to your readout or transmitter with effectively zero energy loss. The 5606 has been proven in common heat transfer fluids such as silicone oil, mineral oil, ethanol, and even liquid nitrogen with no degradation of the transition junction seal or lead wire insulation.

General Specifications
Temperature range
-200 °C to 160 °C
Nominal resistance at 0.01 °C
100 Ω ± 0.1 Ω
Temperature coefficient
0.00385 Ω/Ω/°C
Accuracy[1]
± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C
Short-term repeatability[2]
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Drift[3]
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Hysteresis
± 0.015 °C
Sheath length
50 mm ± 5 mm
(2 in ± 0.2 in)
Sheath diameter
3.1 mm ± 0.1 mm (1/8 in ± 0.004 in)
Sheath material
316 SST
Transition junction temperature range[4]
-200 °C to 160 °C
Transition junction dimensions
No transition
Sensor length
30 mm ± 3mm (1.2 in ±0.1 in)
Sensor location
3 mm ± 1 mm from tip (1.2 in ± 0.04 in)
Minimum insulation resistance
20 MΩ at 23 °C
Minimum immersion length
Full immersion
Maximum immersion depth in liquid medium
Full immersion
Maximum immersion depth in dry Medium
Full immersion
Response time[5]
12 seconds typical
Self heating (in 0 °C bath)
± 0.003 °C
Lead-wire cable type
Enameled copper wire
Lead-wire length
2.4 meters (8 ft)
Lead-wire temperature range
160 °C
Calibration
Calibration not included. NVLAP accredited calibration available. See ordering information.

[1] Includes calibration and 100 hr drift (k = 2).
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[3] After 100 hours at max temp, 95 % confidence (k = 2).
[4] Temperatures outside this range will cause irreparable damage.
[5] Per ASTM E 644. 
 
Calibration Uncertainty
Temperature
-197 °C  1924 Calibration Uncertainty† 0.03 °C
   Tolerance w /out Calibration‡ 0.4 °C
-38 °C  1924 Calibration Uncertainty† 0.03 °C
   Tolerance w /out Calibration‡ 0.2 °C
0 °C  1924 Calibration Uncertainty† 0.03 °C
   Tolerance w /out Calibration‡ 0.1 °C
100 °C  1924 Calibration Uncertainty† 0.045 °C
   Tolerance w /out Calibration‡ 0.3 °C
157 °C  1924 Calibration Uncertainty† 0.05 °C
   Tolerance w/out Calibration‡ 0.4 °C